
PAGE 16 MODEL 110E PUMP OPERATION / MAINTENANCE MANUAL
4.5 PARTS LIST
4.6 CLEAN-UP
The pump fluid cavities may be flushed clean by cycling with the suction (IN) and
discharge (OUT) lines connected to a DI water flushing or rinsing tank. Flushing
using an external pressure source without allowing the pump to cycle will result in
incomplete removal of potentially dangerous chemicals.
NOTE: To effectively eliminate chemical contamination, the pump should cycle
for 10-15 minutes using the flushing DI water as described above.
4.7 DISASSEMBLY
During the life of the pump it will be necessary to perform certain preventative
maintenance procedures to ensure its continued high performance. This section
and the next (4.8 Assembly) are provided for the user’s convenience in
disassembly and re-assembly procedures.
Thoroughly clean / flush the pump using DI water (Refer to Section 4.6
Clean-Up).
Disconnect fluid and air connections from pump.
Unlock pump from quick release base by pulling out locking lever on front of
quick release base. Slide pump forward until it stops. Lift pump off quick
release base.
Remove 3/8” NPT plugs from heads.
Remove tie bolt caps from both heads.
Remove the un-tact flanged nuts from the tie bolt assemblies. Using both
3/8” nut drivers. Leave tie bolts in place.
Komentarze do niniejszej Instrukcji